Coaxial Cable Driver & Matching Amplifiers

FIBER OPTIC TRANSMISSION ALSO AVAILABLE
-
Wideband: DC to over 30MHz
-
Maximum Output: to 10V PK, 20V PK-PK
-
Maximum Input: ±5V to ±200V PK-PK as required. Also to 900kV PK with Ross Voltage Divider
-
Signal Isolation: 500,000 Ohm 450V PK
-
Input Impedance: Up to 2 Megohms
-
Load Impedance: 50 Ohm or greater
-
Power Source: AC or Optional Battery
-
Will drive over 500 Feet of Coaxial Cable
Matching Amplifier
Model VDE-
Wideband: DC to over 30MHz
-
Signal Input
-
0 to ±50 volt peak differential.
-
(0-100V PK-PK) standard. Other available.
-
Peak common mode voltage that will not damage amplifier or distort the output.
-
200 volt steady state
-
450 volt transient
-
-
Impedance
-
1.0 megohm in parallel with 10pF or 2.0 meghom in parallel with 15pF as ordered.
-
-
-
Signal Output
-
0 to ±10 volts (20 volts PK-PK) into 50 ohm or greater load.
-
±200mA. Peak Maximum load current.
-
Matching amplifier isolation of signal output from signal input is 500 kilohm in parallel with 20pF.
-
Gain / frequency. See figure #1 below
-
Undistorted output voltage swing. See figure #2 below
-
Square wave input. Rise and fall time<10ns.
-
Matching amplifier output is calibrated into a specified length of cable and a 50 ohm or greater load as specified by the customer.
-
-
Power Requirements
-
105V to 125V AC, 50/60Hz.
-
Power Consumption-5W.
-
-
Environmental
-
Indoor.
-
0 °C to +50 °C.
-
-
Dimensions
-
3 1/2” high x 7” wide x 8 3/8” deep.
-
-
Miscellaneous
-
Red LED indicates power on.
-
Model VDE-Matching Amplifier includes cabinet, amplifier, 115 volt 50-400Hz power supply, cable and terminations as specified by the customer.
-
Power supply input isolated for 2000 volts.
-
Matching amplifier may also be used with other test equipment such as voltmeters, signal generators, recorders, etc.
-
-
WARNING
-
MATCHING AMPLIFIER OUTPUT COAXIAL CABLE SHIELD IS NOTGROUNDED. FOR SAFETY, OUTPUT COAXIAL CABLE SHIELD SHOULD BE GROUNDED AT COAXIAL CABLE TERMINATING RESISTOR.
-
If output coaxial cable is connected to matching amplifier input ground or matching amplifier 115 volt ground, there may be excessive ground currents causing electrical noise in matching amplifier output.
-
-
CAUTION
-
Do not subject output to reverse voltage.
-
External pickup from coaxial cables can easily exceed safe values if the cables are run near high pulse current conductors.
-
The matching amplifier, coaxial cable and coaxial cable terminating resistor are calibrated as a system.
-
If any of these are changed, the calibration, the high frequency response and the pulse response may change.
-
-
Options On Request
-
Signal Input Variations Available:
-
Any maximum input voltage from ±0.5 to over ±250 volt with proper input range selected, to over 900kV when using Ross Engineering Corp. voltage dividers.
-
-
Signal Output Variations Available:
-
Coaxial cable lengths from 0 to 500 feet
-
-
Power Options Available:
-
Battery or DC operated. 115V, 208V, 240V 50-400Hz.
-
-
Environment Additional Requirements Available:
-
Outdoor
-
Transformer Oil
-
-
Miscellaneous Available:
-
Power on/off switch
-
-

Specifications are for reference only and are subject to change.
Contact Ross Engineering Corp. for current information.